How do I set up a PyroMark CpG Assay?
All relevant information regarding PyroMark CpG Assay setup can be found on the GeneGlobe website. For the Q24 and Q96, the "Sequence to Analyze" and dispensation order should not be copied manually to create a new assay. Instead, the assay file should be downloaded from the web and opened in the PyroMark CpG software, PyroMark Q96 ID v2.5 (or higher) software, and PyroMark Q24 Software to keep important software settings.
When using the PyroMark CpG assays with the PyroMark Q48, use the "Sequence to Analyze" provided in the "product specification" section to create an assay setup file in the PyroMark Q48 software. This is done by selecting New CpG assay and pasting in the Sequence to Analyze (not the "sequence after bisulfite treatment") into the Sequence Before Bisulfite Treatment field and pressing Create Dispensation order.
When using the PyroMark CpG assays with the PyroMark Q48, use the "Sequence to Analyze" provided in the "product specification" section to create an assay setup file in the PyroMark Q48 software. This is done by selecting New CpG assay and pasting in the Sequence to Analyze (not the "sequence after bisulfite treatment") into the Sequence Before Bisulfite Treatment field and pressing Create Dispensation order.