To ensure sensitivity, artus CMV Kits have been optimized to detect low numbers of CMV DNA. The analytical sensitivity of the artus CMV QS-RGQ Kit is 42.5 copies/ml in consideration of the purification and assay setup using the QIAsymphony RGQ (see figure " Sensitive detection of CMV DNA").
Kit | artus CMV RG PCR Kit | artus CMV QS-RGQ Kit |
---|---|---|
Sample type | EDTA plasma | EDTA plasma |
Analytical sensitivity | 57.1 copies/ml | 42.5 copies/ml |
Linear range | n.d. | 79.4 to 1.00 x 108 copies/ml |
artus CMV Kits are based on the amplification and simultaneous detection of a specific region of the CMV genome using real-time PCR. The kits provide high levels of specificity and sensitivity (see figure " Sensitive detection of CMV DNA").
Each artus CMV Kit provides 4 CMV quantitation standards (see figure " Reliable quantitation of CMV load"). Use of the standards enables accurate quantitation of viral load. In addition, the kits contain a second heterologous amplification system to identify possible PCR inhibition. This is detected as an internal control (IC) in a different fluorescence channel from the analytical PCR. The detection limit of the analytical CMV PCR is not reduced.
Kit | artus CMV RG PCR Kit and artus CMV QS-RGQ Kit |
---|---|
Sample type | EDTA plasma |
Amplicon | 105 bp region of the major immediate early (MIE) gene |
artus CMV PCR Kits provide all necessary reagents optimized for reliable CMV DNA detection and quantitation. Simply add template DNA to the ready-to-use PCR master mix and Mg solution, and start the reaction on the appropriate real-time cycler using the optimized cycling program described in the kit handbook.
The temperature profile of the artus CMV RG PCR Kit and the artus CMV QS-RGQ Kit corresponds to the profiles of the artus RG PCR Kits and artus QS-RGQ Kits for BK virus, EBV, HSV-1/2, and VZV. Therefore, these PCR assays can be performed and analyzed in a single run.
The QIAsymphony RGQ workflow solution for CMV detection comprises the QIAsymphony SP for sample preparation, the QIAsymphony AS for assay setup, and the artus CMV QS-RGQ Kit on the Rotor-Gene Q. The system enables reliable pathogen detection with a complete workflow (see figure " Integrated QIAsymphony RGQ system for CMV detection").
For the artus RG and TM PCR Kits, we recommend using the QIAamp DSP Virus Kit. With the artus CMV LC PCR Kit, we recommend using either the QIAamp DSP Virus Kit or the EZ1 DSP Virus Kit.
artus CMV LC, RG, and TM PCR Kits enable rapid and sensitive detection of CMV DNA from EDTA plasma with accurate quantification. Kits are available for use on the ABI PRISM 7000, 7700, and 7900HT SDS, on the LightCycler 1.1/1.2/1.5/2.0 Instruments, or on Rotor-Gene Q instruments.
The artus CMV QS-RGQ Kit is designed to be used with the QIAsymphony RGQ system, providing a complete workflow from sample to CMV DNA detection and quantitation.