How is the Pyromark Q48 cartridge to be cleaned?

Prior to the first run of the day and upon completion of the final run of the day, the injectors are cleaned with high-purity water using the cleaning program of the instrument.

Optionally, a pyrophosphatase clean using the maintenance programs of the instrument software can be performed if needed. 

We recommend the cartridges be exchanged once or twice per quarter and once every second year, depending on throughput.