✓ 24/7 automatic processing of online orders
✓ Knowledgeable and professional Product & Technical Support
✓ Fast and reliable (re)-ordering
Cat. No. / ID: 4506356
✓ 24/7 automatic processing of online orders
✓ Knowledgeable and professional Product & Technical Support
✓ Fast and reliable (re)-ordering
To ensure highest sensitivity, the artus HBV QS-RGQ Kit has been optimized to detect low numbers of HBV DNA. The analytical sensitivity of the artus HBV QS-RGQ Kit is 10.2 IU/ml in consideration of the purification and assay setup using the QIAsymphony RGQ system (see figure " Highly sensitive detection of HBV DNA"). (1 IU/ml corresponds to 8.21 copies/ml for detection of HBV DNA on the Rotor-Gene Q. The conversion factor is an approximation based on an average factor across the assay's dynamic range.)
For highest specificity, validation of the artus HBV QS-RGQ Kit was carried out using various HBV isolates, including all genotypes A–H and related pathogens.
Kit | artus HBV QS-RGQ Kit |
---|---|
Validated sample type | EDTA plasma |
Analytical sensitivity | 10.2 IU/ml |
Linear range | 31.6 to >2 x 107 IU/ml |
Specificity | HBV genomes A–H |
The artus HBV QS-RGQ Kit is based on the amplification and simultaneous detection of a specific region of the HBV genome using real-time PCR. The kit provides high levels of specificity, sensitivity (see figure " Highly sensitive detection of HBV DNA"), and reproducibility over a broad linear range.
Each artus HBV QS-RGQ Kit provides 5 HBV quantitation standards (see figure " Reliable quantitation of HBV load"). Use of the standards enables accurate quantitation of viral load. In addition, the kit contains a second heterologous amplification system to identify possible PCR inhibition. This is detected as an internal control (IC) in a different fluorescence channel from the analytical PCR. The detection limit of the analytical HBV PCR is not reduced.
Kit | artus HBV QS-RGQ Kit |
---|---|
Validated sample type | EDTA plasma |
Amplicon | 134 bp core region of the HBV genome |
The artus HBV QS-RGQ Kit provides all necessary reagents optimized for reliable HBV DNA detection and quantitation. Simply add template DNA to the ready-to-use PCR master mix and start the reaction on the appropriate real-time cycler using the optimized cycling program described in the kit handbook.
The QIAsymphony RGQ workflow solution for HBV detection comprises the QIAsymphony SP for sample preparation, the QIAsymphony AS for assay setup, and the artus HBV QS-RGQ Kit on the Rotor-Gene Q. The system enables reliable pathogen detection with a complete IVD workflow (see figure " Integrated QIAsymphony RGQ system for HBV detection").
The artus HBV QS-RGQ Kit is designed to be used with the QIAsymphony RGQ system, providing a complete IVD workflow from sample to HBV DNA detection and quantitation.