The STAR Q Punch AS Instrument automates sample punching and STR assay setup from human identity reference samples such as database samples and paternity samples on collection cards. The instrument works with two collection card types and has pre-validated protocols for use with Investigator 24plex GO! Kits, Investigator ESSplex SE GO! Kits and Investigator IDplex GO! Kits. The instrument is designed for use with QIAGEN EasiCollect cards, Bode Buccal Collectors and Copan NUCLEIcard system collection cards. The STAR Q Punch AS combines QIAGEN’s expertise in performance and process safety with high-throughput liquid handling to enable processing of up to 360 samples per day with no compromise in quality.
The STAR Q Punch AS is a high-throughput punching and liquid handling platform designed for assay setup for downstream PCR analysis of human identity samples. The system is designed for processing samples collected on QIAGEN EasiCollect cards, Bode Buccal Collectors and Copan NUCLEIcard collection devices.
QIAGEN’s Investigator STR GO! Kits for direct amplification of reference samples can directly be processed on the instrument. This means that the results include QIAGEN's unique Quality Sensor, so any sample that fails to give optimal results the first time can be reprocessed with the most appropriate rework strategy, maximizing success rates and minimizing costs for your laboratory.
Unlike most other high-throughput punching and liquid handling systems, the STAR Q Punch AS comes with pre-validated protocols for punching and assay setup, meaning implementation is kept to a minimum.
The STAR Q Punch AS comes ready to run, with all protocols preinstalled and fully validated. Our experienced application specialists will manage instrument installation and training and ensure the system is quickly up and running.
Sample cards, reagents and consumables are loaded onto the instrument, which performs safety checks and bar code scans. The system then automates sample visualization and punching from sample collection cards directly into a 96-well PCR plate before proceeding with STR PCR setup according to predefined protocols optimized for use with Investigator STR GO! Kits.
All protocols and other system functions are selected and controlled using a simple-to-use Graphical User Interface (GUI).
Features | Specifications |
---|---|
Voltage | 100 VAC / 115 VAC / 230 VAC |
Dimensions | Width: 1124 mm, Height: 903 mm, Depth: (with Autoload) 1006 mm |
Modal precision | x-y-z positional accuracy of 0.1 mm |
Power consumption | Standby power consumption: 100 VA Maximum power consumption: 600–1000 VA, depending on configuration |
Delayed action fuse – 600 VA | 115 V: 6.3 A (T6.3AL250) 230 V: 3.15 A (T3.15AL250) |
Noise level | <65 dBA (regarding EN27779) <46 dBA in standby mode |
Channels | 4 channels |
Tip sizes | Intermediate volume 50 µl |
Heat | The power consumed will be transferred to head. E.g., 600 or 1000 watts of heat = 600 or 1000 joules/second |
Temperature range | 15–35°C |
Frequency | 50/60 Hz ± 3% |
Installation category | II |
Work area dimensions | Width (x): 675 mm, Height (z): 145 mm, Depth (y): 497 mm |
Maximum power consumption | 600 VA or 1000 VA (depending on configuration) |
Delayed action fuse – 1000 VA | 115 V: 10 A (T10AL250) 230 V: 5 A (T5AL250) |
Pollution degree | 2 |
Weight | 135 kg |
Recommended computer | Intel Core 2 Duo, ≥ 4 GB RAM, 500 GB hard drive, 16x DVD +/-RW, DirectX 250 MB graphic card, MS Windows 7 Professional (not included in shipment) |
Relative humidity | 30%–85% (non-condensing, indoors) |
Communication | USB or RS232 with dual processor board Ethernet or USB with LAN dual processor board |
Storage conditions | Relative humidity: 10–90% (non-condensing, indoors) |
Use | Indoor use only |
Materials used in instrument and accessories | Steel: 1.4301, 1.4305, 1.4310, 1.4435 Aluminum: AA 5083.0, various anodizing Bronze Brass Brazing solder: Zinc Plastics: PE, PP, PTFE, PEEK, FFPM, EPT, EPDM, FPM, NBR, POM, PCDF, FFKM, Hypalon, SI, Kapton, Polyamide PCBs |
Altitude | Up to 2000 m (6500 ft) |